Main Page → Bertec Force Plate Setup
Scope
This page provides instructions on how to integrate a Bertec force plate system with an OptiTrack motion capture system.
Overview
When a motion capture system is used in conjunction with force plates, they work together as an efficient tool for various research applications including biomechanical analysis, clinical gait analysis, physiology research, sports performance research, and many more. An OptiTrack motion capture system can synchronize with force plates to obtain both kinematic and kinetic measurements. Note that force plate integration is supported only with a Prime camera system using the eSync synchronization hub. This page provides quick guidelines for setting up and configuring force plates — with digital outputs — along with the OptiTrack motion capture system.
For detailed information on specifications and configurations on the force plates, refer to the documentation provided by the force plate manufacturer.
Analog Platforms
Requirements
Hot plugging is not supported with the integration. When a new device is connected to the system, you must re-start Motive to instantiate it.
In order to integrate force plate systems with Motive, you will need to setup the required drivers and plugins. Motive installer is packaged with the Peripheral Device module which can be added. During the Motive installation, a list of program features will be shown in the Custom Setup section. Here, change the setting for the Peripheral Device module, as shown in the below image, so that the module is installed along with Motive Files.
Note : Even if you are not using NI-DAQ, it is still necessary to install NI-DAQmx drivers that come up next in the installer.
Note: You cannot run Digital Acquire and Motive together and have them connected to the plate(s) at the same time.
If the hardware and software for the force plates are configured and successfully recognized, Motive will list out the detected force plates with number labels (1, 2, etc..). Motive will notify you of incorrect or nonexistent force plate calibration files. When the devices are properly instantiated in Motive, the following status messages will be shown under that Status Log.
Calibrate the capture volume as normal to get the orientation of the cameras (see the Quick Start Guide or Calibration page for more information). The position of the force plate is about the center of the volume, and when you recalibrate or reset the ground plane, you will need to also realign the position of your force plates for best results.
On the CS-400 calibration square, pull the force plate alignment tabs out and put the force plate leveling jigs at the bottom. The leveling jigs align the calibration square to the surface of your force plate. The alignment tabs allow you to put the CS-400 flush against the sides of your force plate giving the most accurate alignment.
Place the calibration wand on the force plate so that vertex of the wand is located at the right-hand corner of the side where the cable input is located (as shown in the image below). A correct placement of the calibration square is important because it determines the orientation of the force plate and its local coordinate axis within the global system. The coordinate systems for force plates are independent of the system used Motive.
After placing the calibration square on the force plate, select the CS-400 markers in Motive. Right click on the force plate you want to locate, and click Set Position. When there are multiple force plates in a volume, you may need to step on the force plate to find which platform the calibration square is on. In Motive, uncalibrated force plates will light up in green and a force vector will appear when you step on the plate. Repeat step 4 and 5 for other force plates as necessary.
Referencing to the markers on the calibration square, Motive defines the location of the force plate axis within the global coordinate system. When setting the position, the previously defined force plate dimension values and mechanical-to-electrical center offset values will be applied and reflected in the Perspective View pane.
Tip: To double check that the dimensions are modified properly, you can place extra retroreflective markers on each corner of the platform and monitor the coincidence of the markers position with the force plate assets from the perspective view.
If the force plate dimensions are not automatically configured, you need to enter the dimensions of the force plate in the Devices pane after calibrating its positions. Go to the devices pane and select the force plate, under its force plate properties, enter the length and width (in meters) values for the corresponding plates as reported in the specifications.
After you have calibrated each of your force plates, remove the CS-400 from the volume. Right click one of your force plates in Motive and click Zero (all). This will tare the scale and set the current force on the plate data to 0. This will account for a small constant amount of measurement offset from the force plate. Remember that it zeros all of the force plates at once. So make sure there are no objects on any of the force plates.
Note: Zeroed scales of Bertec force plates are saved within their software driver, and each time the driver restarts, these settings are refreshed. This means that the force plate zero setting will be refreshed each time you start Motive, or each time the device is disabled and enabled back again in Motive. Please be aware of this behavior and zero your plates when necessary. In Motive, there is a Zero On Enable property setting for Bertec force plates under the Devices pane, and enabling this setting will automatically zero your plate each time the device is enabled or when Motive restarts. The Zero On Enable setting is enabled by default.
It is best to configure the sampling rate after the sync configurations have been set up which will be covered in the following section. Sampling rate of force plates is configured through the devices pane, by modifying the framerate-multiplier (e.g. 10.0x) next to the sampling rate. Configuring the multiplier ensures that sampling rate of the force plate is always a whole multiple of the camera framerate. Note that Bertec force plates support the 1000 Hz sampling rate only.
When synchronizing through the eSync, use the following steps for configuring the sync settings in Motive. This will allow both systems to be simultaneously triggered with reference to the master synchronization device. The Recording Gate signal must be sent out from the eSync and inputted into the Bertec force plate system for synchronization.
IMPORTANT NOTE: For this synchronization setup to work properly, the Bertec amplifier firmware must be updated to its most recent version (above June 2016 release). Contact Bertec for instructions on updating the firmware.
Note: Bertec digital force plates can collect data at a1000 Hz sampling rate only.
Before you start recording, you may want to validate that the camera and force plate data are in sync. There are some tests you can do to examine this.
The first method is to record dropping a retroreflective ball/marker onto the platform few times. The bouncing ball produces a sharp transition when it hits the surface of the platform, and it makes the data more obvious for validating the synchronization. Alternately, you can attach a marker on a tip of the foot and step on and off the force plate. Make sure that your toe — closest to the marker — strikes the platform first, otherwise the data will seem off even when it is not. You can then monitor the precise timing of the ball or the foot impacting the force plate and compare them between the mocap data and the force plate data. ↑
The following is an example of validating good synchronization outcomes:
Above configurations set the sync chain to be in Free Run Mode. This means that the recording of both the mocap system and the force plate system are triggered simultaneously at the same time and each system runs at its own rate.
Two systems, however, are synchronized at the trigger point but not at every frame. For this reason, alignment of the mocap data and the force plate data may gradually drift from each other for longer captures. But this is not a problem since the sync chain will always be re-synchronized each time recording in Motive is triggered. Furthermore, Takes in general do not last too long for this drift to take effect on the data.
However, this could be an issue when live-streaming the data since recording is never initiated and two systems will be synchronized only when Motive first launches. To zero out the drift, the ReSynch feature can be used. Right-click on force plates from either the Devices pane or the perspective view, and select Resynch from the context menu to realign the sampling timing of both systems.
First of all, it is important to note that without a master synchronization device, recording for the camera system and the force plates will not be triggered exactly at the same time. If you want your system to be timed perfectly, we recommend synchronizing through the eSync 2.
If you are using an Ethernet system without an eSync, ignore the synchronization configuration options. From the Devices pane, make sure that that the force plate sampling rate is a multiple of the tracking frames per second (e.g. 100 Hz and 1000 Hz). Motive will alert you if the sampling rates of cameras and force plates disagree. When the force plate sampling rate is not set to an integer multiple of the camera frame rate, force plate data may record improperly and cause an error.
Note: Bertec digital force plates can collect data up to a maximum of 1000 Hz sampling rate.
Starting from Motive 2.0, configurations on external devices (e.g. Force Plates and NI-DAQ devices) are persisted between different launches of Motive.
All of the configured device settings, including the calibration, get saved on Device Profile XML files. When you exit out of Motive, updated device profiles will be saved under the program data directory (C:\ProgramData\OptiTrack\Motive\DeviceProfiles), and this file gets loaded again when you restart Motive. The persistent settings folder can be accessed through Help → Application Folders → Persistent Setting. This XML file ensures that all of the device settings are persisted each time you close and restart Motive.
C:\ProgramData\OptiTrack\Motive\DeviceProfiles
Force plate data can be monitored from the Graph View pane. You will need to configure a custom graph layouts to show force plate data. As shown in the images, make sure the desired force plate data channels (Fx, Fy, Fz, Mx, My, or Mz) are selected to be plotted. Then, when you select a force plate in Motive, and the data from the corresponding channels will be plotted on the graphs. When both reconstructed markers and force plate channels are selected, the force plot will be sub-sampled in order to be plotted along with trajectory data. For more information about how to configure graph layouts, read through the Graph View pane page.
Notes
We recommend the following programs for analyzing exported data in biomechanics applications:
Since Motive uses a different coordinate system than the system used in common biomechanics applications, it is necessary to modify the coordinate axis to a compatible convention in the C3D exporter settings. For biomechanics applications using z-up right-handed convention (e.g. Visual3D), the following changes must be made under the custom axis.
This will convert the coordinate axis of the exported data so that the x-axis represents the anteroposterior axis (left/right), the y-axis represents the mediolateral axis (front/back), and the z-axis represents the longitudinal axis (up/down).
Force plate data and the tracking data can be exported into CSV files as well. When a Take file is exported into a CSV file. Separate CSV files will be saved for each force plate and it will contain the force, moment, and center of pressure data. Exported CSV file can be imported for analysis.
To stream tracking data along with the force plate data, open the Data Streaming Pane and check the Broadcast Frame Data, and make sure that you are not streaming over the camera network. Read more about streaming from the Data Streaming workflow page.
Motive can stream the tracking data and the force plate data into various applications — including Matlab — using NatNet Streaming protocol. Find more about NatNet streaming from the User's Guide included in the download.