AMTI Force Plate Setup

Main PageAMTI Force Plate Setup

Scope

This page provides instructions on integrating an AMTI Force plate system with an OptiTrack motion capture system.

Overview

When a motion capture system is used in conjunction with force plates, they work together as an efficient tool for various research applications including biomechanical analysis, clinical gait analysis, physiology research, sports performance research, and many more. An OptiTrack motion capture system can synchronize with force plates to obtain both kinematic and kinetic measurements. Note that force plate integration is supported only with a Prime camera system using the eSync synchronization hub. This page provides quick guidelines for setting up and configuring force plates — with digital outputs — along with the OptiTrack motion capture system.

For detailed information on specifications and configurations on the force plates, refer to the documentation provided by the force plate manufacturer.

Info2.png

Analog Platforms

Analog force plate devices can only be implemented via DAQ devices. Incoming voltage signals can be detected through the data acquisition channels, but force plate related software features (vectors, position calibration, etc.) will not be supported in Motive for the analog platforms. Refer to the NI-DAQ Setup page for detailed instructions on integrating analog devices.

Required Components

  • Supported Amplifier Models: AMTI Gen 5, AMTI Optima.
  • Force platforms that are compatible with the above amplifier models.
  • Prime series Ethernet camera system with the eSync synchronization hub.
  • Motive 1.9 or above.


Hardware Setup


Component Wiring (AMTI)

AMTI Force Plate System Setup

Connect each force plate into the host PC. For force plate systems with external amplifiers, the platform must be connected to the amplifier which uplinks to the host computer. For detailed instructions on setting up the Force Plate system with a host PC, refer to the AMTI documentation.

Camera System Setup

Setup the OptiTrack camera system and place the force plate(s) at the desired location(s); ideally, near the center of the volume. See Quick Start Guide or Hardware Setup page for details.

Wiring the eSync with the Gen 5 Amplifier

For accurate synchronizations, the eSync 2 synchronization hub must be used. The eSync 2 has signal output ports that are used to send out synchronization signals to slave devices. Connect the BNC output ports of the eSync to sync input ports (Genlock/Trigger Input) of force plate amplifiers.
If force plate systems have RCA sync ports, use RCA cables along with the 50 Ohm BNC Male to 75 Ohm RCA Jack Adapters included with the eSync 2 to connect the amplifiers. The above wiring diagram shows how force plate systems need to be connected to an Ethernet camera system through the eSync 2.

Multiple Devices Sync

There are total four output ports on the eSync 2, and multiple force plates and external devices can be integrated if needed. Consult our Engineers for multiple force plate synchronizations.


  • Diagram for Prime Camera System with integrated force plates.
    Diagram for Prime Camera System with integrated force plates. (*) Some force plates don't have external amplifiers, but instead, have their amplifiers integrated within the platform. In this case, connect sync cables and the USB cables directly to the host PC.
  • The eSync 2 output and input ports
  • Software Setup


    AMTI Software Setup

    Before setting up the force plates in Motive, make sure software components required by the force plate system is installed on the computer. AMTI's software (e.g. AMTINetForce) must be able to detect and initialize the connected devices in order for the force plates to be properly initialized and used in Motive. Once this has been confirmed working, start setting up Motive. Please refer to manufacturer documentation for more information.

    Optima amplifier detected in the NetForce software.

    Peripheral Device Module

    In order to integrate force plate systems with Motive, you will need to setup the required drivers and plugins. Motive installer is packaged with the Peripheral Device module which can be added. During the Motive installation, a list of program features will be shown in the Custom Setup section. Here, change the setting for the Peripheral Device module, as shown in the below image, so that the module is installed along with Motive Files.

    Info2.png

    Note : Even if you are not using NI-DAQ, it is still necessary to install NI-DAQmx drivers that come up next in the installer.

    Warning2.png

    Installation Note: For integration into Motive, the NI-DAQmx 15.1.1 or later runtime driver must be installed. If you are already using an older version of the NI-DAQmx runtime and Motive is having problems recognizing the connected device, update the driver or uninstall and re-install the packaged version of the driver before contacting Support. In Motive, you can inspect device connection status via the Status Log panel which can be accessed under the View tab in Motive.

    Warning2.png

    Notes on NI-DAQmx version 19.x: There is a known issue with the latest NI-DAQmx version (19.x) where it crashes Motive on startup. If you have this version of the driver installed, please reinstall with the recommended version (15.1.1) provided with the Motive installer.


    Installing OptiTrack Peripheral Module. DAQmxInstall.gif


    Force Plate Setup in Motive

    1. Start Motive

    If the hardware and software for the force plates are configured and successfully recognized, Motive will list out the detected force plates with number labels (1, 2, etc..). Motive will notify you of incorrect or nonexistent force plate calibration files. When the devices are successfully instantiated in Motive, the Status Log will indicate that the device has been created and loaded.

    • Initial view in Motive after setting up force plates.
      Motive with force plates.

    2. Calibrate Cameras

    Calibrate the capture volume as normal to get the orientation of the cameras (see the Quick Start Guide or Calibration page for more information). The position of the force plate is about the center of the volume, and when you recalibrate or reset the ground plane, you will need to also realign the position of your force plates for best results.

    Motive after setting up force plates and calibration.
    Motive with force plates and camera calibration.


    3. Setup CS-400

    On the CS-400 calibration square, pull the force plate alignment tabs out and put the force plate leveling jigs at the bottom. The leveling jigs align the calibration square to the surface of your force plate. The alignment tabs allow you to put the CS-400 flush against the sides of your force plate giving the most accurate alignment.

    CS-400 calibration square with force plate force plate parts.
    CS-400 calibration square with force plate force plate parts.


    4. Place CS-400 on force plate

    Place the calibration wand on the force plate so that vertex of the wand is located at the right-hand corner of the side where the cable input is located (as shown in the image below). A correct placement of the calibration square is important because it determines the orientation of the force plate and its local coordinate axis within the global system. The coordinate systems for force plates are independent of the system used Motive.

    Info2.png

    AMTI Force Plates

    AMTI force plates use the right-hand system. The long arm of CS-400 will define the Y axis, and the short arm will define the X axis of the force plate. Accordingly, Z axis is directed downwards for measuring the vertical force.

  • Force plate with CS-400 aligned properly.
    Force plate with CS-400 aligned properly.
  • Calibrated force plate position and orientation. X and Y axis is shown.
  • 5. Set force plate position in Motive.

    After placing the calibration square on the force plate, select the CS-400 markers in Motive. Right click on the force plate you want to locate, and click Set Position. When there are multiple force plates in a volume, you may need to step on the force plate to find which platform the calibration square is on. In Motive, uncalibrated force plates will light up in green and a force vector will appear when you step on the plate. Repeat step 4 and 5 for other force plates as necessary.
    Referencing to the markers on the calibration square, Motive defines the location of the force plate coordinate system within the global coordinate system.
    Motive uses manufacturer defined X, Y, and Z mechanical-to-electrical center offset when calculating the force vector and the center of pressure. For digital based plates, this information is available from the SDK and also stored in the plate's on-board calibration data.
    Setting the position of a force plate in Motive.
    Setting the position of a force plate in Motive. The number label on the force plate is inverted because the force plate position and orientation has not been calibrated yet.

    6. Zero force plates.

    After you have calibrated each of your force plates, remove the CS-400 from the volume. Right click one of your force plates in Motive and click Zero (all). This will tare the scale and set the current force on the plate data to 0. This will account for a small constant amount of measurement offset from the force plate. Remember that it zeros all of the force plates at once. So make sure there are no objects on any of the force plates.

    Set the force plate data to zero for more accurate data.
    Set the force plate data to zero for more accurate data.

    7. Set Sampling Rate

    Configuring force plate sampling rate from Devices pane.

    Sampling rates of force plates can be set from the Devices pane. For collected data from two systems to be aligned, the sampling rate of force plates must be a whole multiple of the camera system's framerate. By adjusting the Rate Multiplier values in the Devices pane, sampling rates of the force plates can be modified. First, pick a frame rate of the camera system and then adjust the rate multiplier values to set force plates to the desired sampling rate.

    Please note that only specific sampling rates may be supported depending on the amplifier models.

    Info2.png

    Supported force plate sampling rates:

    For AMTI force plates support the following sampling rates depending on the amplifier used. For the most up-to-date information, consult their documentation. The supported sampling rates (Hz) are the following:

    • AMTI Gen 5 Amplifier: 2000, 1800, 1500, 1200, 1000, 900, 800, 600, 500, 450, 400, 360, 300, etc...
    • AMTI Optima Amplifier: 1000, 600, 500, 300, 250, 200, 150, 125, 120, 100, 60, 50, 30, 25, 15, 10.

    Synchronization Configuration

    When synchronizing through the eSync, use the following steps to configure the sync settings in Motive. This will allow both systems to be triggered simultaneously with reference to the master synchronization device, the eSync.

    Example settings in the cameras and synchronization panes for eSync.
    Example settings in the cameras and synchronization panes for eSync.

    Sync Configuration Steps: eSync & AMTI Force Plate

    1. Open the Devices pane and the Synchronization pane.
    2. In the Synchronization pane, select the Custom Synchronization mode from the dropdown menu.
    3. Set the Sync Input → Source to Internal Clock.
    4. Set the Sync Input → Clock Freq to 1000 Hz, and set the Sync Input → Input divider to 10.
    5. Set the Sync Output 1-4 → Type to Record Start/Stop Pulse or Recording Gate triggered sync.
    6. Click Apply from the Synchronization pane.
    7. In the Devices pane, select the force plates group and double check that the Sync Source is set to Free Run for the force plates.

    How to Validate your Synchronization

    Before you start recording, you may want to validate that the camera and force plate data are in sync. There are some tests you can do to examine this.

    The first method is to record dropping a retroreflective ball/marker onto the platform few times. The bouncing ball produces a sharp transition when it hits the surface of the platform, and it makes the data more obvious for validating the synchronization. Alternately, you can attach a marker on a tip of the foot and step on and off the force plate. Make sure that your toe — closest to the marker — strikes the platform first, otherwise the data will seem off even when it is not. You can then monitor the precise timing of the ball or the foot impacting the force plate and compare them between the mocap data and the force plate data.

    The following is an example of validating good synchronization outcomes:

    Good synchronization


    ReSynch

    Re-aligning initial sampling timing of the force plate.
    When two systems are synchronized by recording trigger signals (Recording Gate or Recording Pulse), both systems are in Free Run Mode. This means that the recording of both the mocap system and the force plate system are triggered simultaneously at the same time and each system runs at its own rate.
    Two systems, however, are synchronized at the recording trigger but not by per frame basis. For this reason, alignment of the mocap data and the force plate data may gradually drift from each other for longer captures. But this is not a problem since the sync chain will always be re-synchronized each time recording in Motive is triggered. Furthermore, Takes in general do not last too long for this drift to take effect on the data.
    However, this could be an issue when live-streaming the data since recording is never initiated and two systems will be synchronized only when Motive first launches. To zero out the drift, the ReSynch feature can be used. Right-click on force plates from either the Devices pane or the perspective view, and select Resynch from the context menu to realign the sampling timing of both systems.

    }}}} </tab> </tabs>


    Ethernet System Force Plate without eSync Configuration

    If you are using an Ethernet system without an eSync, ignore the synchronization configuration options. From the Devices pane, make sure that that the force plate sampling rate is a multiple of the tracking frames per second (fps), ex 100fps and 1000fps. Motive will alert you if the sampling rates for cameras and force plates disagree.

    Without a master synchronization device, recording for the camera system and the force plates will not be triggered exactly at the same time. This can cause problems for longer takes because the sampling timing for mocap data and the force plate data will eventually deviate from each other. If you want your system to be timed perfectly, we recommend synchronizing through the eSync 2.

    When the force plate sampling rate is not set to an integer multiple of the camera frame rate, force plate data may record improperly; which is shown in the following screenshot.

    Graph of bad data.
    The force plate stops recording when its sampling rate disagrees with the frame rate of the camera system.

    Force Plate Data in Motive


    To view and confirm the live force plate data, open the editor from the Timeline Pane. Then, open the Project Pane (or Devices pane) and select one of the force plates. A list of available channels will appear under each force plate instance. Here, you can confirm that your force plates are working properly. Select Fx, Fy, Fz, Mx, My, or Mz channels to view the live force plate data from the timeline. Multiple channels can be displayed at once. When both reconstructed markers and force plate channels are selected, the force plot will be sub-sampled in order to be plotted along with trajectory data.

    If you wish to double check the force plate position calibration, you may create a long trackable rigid body and use it to apply force against force plate. If the force plate location is precisely calibrated in Motive, the force vector will go right through the rigid body.

    • Graph of live force plate data.
      Graph of live force plate data.
    • A resultant force vector displayed in Motive.

    Info2.png

    Notes

    • The force and moment data reflects the coordinate system defined by the force plate manufacturer, which is typically the Z-down right-handed coordinate system. Note: This convention is independent of the global coordinate system used in Motive. Thus, the Fz components represent the vertical force. For more in-depth information, refer to the force plate specifications.

    Data Export


    The MotionMonitor biomechanics analysis software.
    Visual3D biomechanics analysis software provided by C-Motion

    We recommend the following programs for analyzing exported data in biomechanics applications:

    C3D Export

    Motive exports tracking data and force plate data into C3D files. Exported C3D files can then be imported into a biomechanics analysis and visualization software for further processing. See the Data Export or Data Export: C3D page for more information about C3D export in Motive. Note that the coordinate system used in Motive (y-up right-handed) may be different from the convention used in the biomechanics analysis software.

    C3D Axes

    Common Conventions

    C3D export setting for applications using z-up right-handed coordinate systems.

    Since Motive uses a different coordinate system than the system used in common biomechanics applications, it is necessary to modify the coordinate axis to a compatible convention in the C3D exporter settings. For biomechanics applications using z-up right-handed convention (e.g. Visual3D), the following changes must be made under the custom axis.

    • X axis in Motive should be configured to positive X
    • Y axis in Motive should be configured to negative Z
    • Z axis in Motive should be configured to positive Y.

    This will convert the coordinate axis of the exported data so that the x-axis represents the anteroposterior axis (left/right), the y-axis represents the mediolateral axis (front/back), and the z-axis represents the longitudinal axis (up/down).

    CSV Export

    Force plate data and the tracking data can be exported into CSV files as well. When a Take file is exported into a CSV file. Separate CSV files will be saved for each force plate and it will contain the force, moment, and center of pressure data. Exported CSV file can be imported for analysis.

    Data Streaming

    To stream tracking data along with the force plate data, open the Data Streaming Pane and check the Broadcast Frame Data, and make sure that you are not streaming over the camera network. Read more about streaming from the Data Streaming workflow page.

    Motive can stream the tracking data and the force plate data into various applications — including Matlab — using NatNet Streaming protocol. Find more about NatNet streaming from the User's Guide included in the download.

    Warning2.png

    Number of Force Plates

    At the time of writing, there is a hard limit on the maximum number of force plate data that can be streamed out from Motive. Please note that only up to 8 force plate data can be streamed out from Motive and received by a NatNet application.


    Troubleshooting


    1) Received force plate data is intermittent.

    • If the force plate is using a 15 feet USB cable, try a shorter cable with a length of maybe 6 or 10 feet. Long USB cables may cause device disconnections.